Rohde and Schwarz FPC1000 5kHz – 3GHz Spectrum Analyzer

$3,312.00

Technical Specs

Rohde and Schwarz FPC1000 - 5kHz - 3GHz Spectrum Analyzer (FPC-P3, order # 1328.6660P03) Specs

Frequency range R&S®FPC1000
with R&S®FPC-B2 option 5 kHz to 2 GHz
with R&S®FPC-B3 option 5 kHz to 3 GHz
Frequency resolution 1 Hz
Aging per year 1 × 10–6
Temperature drift 0 °C to +30 °C
+30 °C to +50 °C 3 × 10–6
Achievable initial calibration accuracy 5 × 10–7
Total reference uncertainty (time since last adjustment × aging rate) + temperature drift + calibration accuracy
Marker resolution 0.1 Hz
Uncertainty ±(marker frequency × reference uncertainty + 10 % × resolution bandwidth + ½ (span / (sweep points – 1)) + 1 Hz)
Number of sweep (trace) points 1183
Marker tuning frequency step size span/1182
Frequency counter resolution 0.1 Hz
Count uncertainty SNR > 25 dB
Frequency span 0 Hz, 10 Hz to 1 GHz
with R&S®FPC-B2 option 0 Hz, 10 Hz to 2 GHz
with R&S®FPC-B3 option 0 Hz, 10 Hz to 3 GHz
Span uncertainty 1 % (nom.)
Spectral purity SSB phase noise f = 500 MHz
Carrier offset 30 kHz
100 kHz < –98 dBc (1 Hz), –103 dBc (1 Hz) (typ.)
1 MHz < –120 dBc (1 Hz), –125 dBc (1 Hz) (typ.)
Sweep time span = 0 Hz
10 Hz ≤ span ≤ 600 MHz 10 ms to 1000 s
span > 600 MHz 10 ms × span/600 MHz to 1000 s
Uncertainty span = 0 Hz
span ≥ 10 Hz 3 % (nom.)
Range -3 dB bandwidth
Bandwidth accuracy 1 Hz ≤ RBW ≤ 300 kHz
300 kHz < RBW ≤ 1 MHz < 10 % (nom.)
Selectivity 60 dB:3 dB < 5 (nom.) (Gaussian type filters)
Range -3 dB bandwidth
Display range displayed noise floor to +30 dBm
DC voltage 50 V
CW RF power 33 dBm (= 2 W)
Peak RF power duration < 3 s
Max. pulse voltage 150 V
Max. pulse energy pulse width 10 μs
Third-order intercept (TOI) intermodulation-free dynamic range, signal level 2 × –20 dBm, RF attenuation = 0 dB, RF preamplifier = off
fin = 1 GHz +7 dBm (meas.)
fin = 2.4 GHz +10 dBm (meas.)
Second harmonic intercept (SHI) RF attenuation = 0 dB, RF preamplifier = off, signal level = –40 dBm
fin = 20 MHz to 1.5 GHz –60 dBc (nom.)
Displayed average noise level dB RF attenuation, termination 50 Ω, RBW = 100 Hz, VBW = 10 Hz, sample detector, log scaling, normalized to 1 Hz
1 MHz to 10 MHz < –127 dBm,–135 dBm (typ.)
10 MHz to 2 GHz < –142 dBm, –150 dBm (typ.)
2 GHz to 3 GHz < –138 dBm, –147 dBm (typ.)
1 MHz to 10 MHz < –147 dBm, –157 dBm (typ.)
10 MHz to 2 GHz < –158 dBm, –165 dBm (typ.)
2 GHz to 3 GHz < –155 dBm, –163 dBm (typ.)
Image frequencies fin – 2 × 30.15 MHz
fin – 2 × 830.15 MHz < –65 dBc (nom.)
fin – 2 × 4042.65 MHz –60 dBc (nom.)
Intermediate frequencies 30.25 MHz, 830.25 MHz, 4042.65 MHz
Other interfering signals, signal level – RF attenuation < –30 dBm spurious at fin – 2021.325 MHz
Other interfering signals, related to local oscillators Δf ≥ 300 kHz
Residual spurious response input matched with 50 Ω, without input signal, RBW ≤ 30 kHz, f ≥ 3 MHz, RF attenuation = 0 dB, Wi-Fi function disabled
Logarithmic level axis 1/2/5/10/20/50/100 dB, 10 divisions
Linear level axis 0 % to 100 %, 10 divisions
Number of traces 2
Trace detectors max. peak, min. peak, auto peak, sample, RMS
Trace functions clear/write, max. hold, min. hold, average, view
Setting range of reference level -130 dBm to +30 dBm
Units of level axis dBm, dBmV, dBμV, V, W
Absolute level uncertainty at 100 MHz +20 °C to +30 °C
Frequency response (+20 °C to +30 °C) 100 kHz ≤ f < 10 MHz
10 MHz ≤ f ≤ 3 GHz < 1 dB
Attenuator uncertainty < 0.3 dB
Uncertainty of reference level setting < 0.1 dB (nom.)
Display nonlinearity SNR > 16 dB, 0 dB to –50 dB, logarithmic level display
Bandwidth switching uncertainty reference: RBW = 10 kHz
Total measurement uncertainty 95 % confidence level, +20 °C to +30 °C, SNR > 16 dB, 0 dB to –50 dB below reference level, RF attenuation auto
10 MHz ≤ f ≤ 3 GHz < 1.25 dB, 0.5 dB (typ.)
Trigger source free run, video, external
External trigger level threshold low → high transition
high → low transition 0.7 V
maximum 3.0 V
Impedance 50 Ω (nom.)
Connector N female
VSWR 5 kHz ≤ f ≤ 1 GHz
1 GHz < f ≤ 3 GHz < 2 (nom.)
Input attenuator RF input only
AF demodulation types AM and FM
Connector 3.5 mm mini jack
Output impedance 32 Ω (nom.)
Voltage (open circuit) VRMS adjustable from 0 V to > 100 mV
Connector BNC, 50 Ω
Mode ext. reference, ext. trigger
External reference required level
Frequency 10 MHz
External trigger threshold low → high transition
high → low transition 0.7 V
AC supply input specifications
Power consumption 14 W (nom.)
Safety IEC 61010-1, EN 61010-1, UL 61010-1, CAN/CSA-C22.2 No. 61010.1
Test mark VDE, GS, CSA
Languages Chinese, English, French, German, Italian, Hungarian, Japanese, Korean, Portuguese, Russian, Spanish
Command set SCPI 1997.0
LAN interface 10/100BASE-T, RJ-45
USB type B plug, version 2.0
Size 10.1"
Resolution 1366 × 768 pixel
Pixel errors < 2 pixel
Speaker internal
USB interface type A plug, version 2.0
number of interfaces 2
Mass memory memory stick (not supplied), size ≤ 4 Gbyte, USB version 1.1 or 2.0
Data storage internal
on memory stick, ≥ 1 Gbyte > 5000 instrument settings and traces
Temperature operating temperature range
storage temperature range -20 °C to +70 °C
Climatic loading relative humidity
Vibration sinusoidal
random EN 60068-2-64
Shock 40 g shock spectrum, in line with MIL-STD-810F, method 516.4 procedure 1, EN 60068-2-27
EMC in line with European EMC Directive 2004/108/EC including   CISPR 11/EN 55011/group 1   class A (emission)   EN 61326 table 2 (immunity, industrial)
Dimensions (W × H × D) without feet
including feet 396 mm × 185 mm × 156 mm   (15.6 in × 7.3 in × 6.1 in)
Weight 3 kg (6.61 lb)
Recommended calibration interval 1 year
Reviews (0)

Reviews

There are no reviews yet.

Be the first to review “Rohde and Schwarz FPC1000 5kHz – 3GHz Spectrum Analyzer”
SKU: ROHDE-&-SCHWARZ-FPC-P3 Category: Brand: