Temperature range at 23 °C –95 °C to 140 °C (–139 °F to 284 °F)
Display  accuracy ± 0.2 °C full range
Accuracy with  external reference [3] ± 0.05 °C full range
Stability ± 0.015 °C full range
Axial uniformity at 40 mm (1.6  in) ± 0.05 °C full range
Radial gradient ±0.01 °C full range
Loading effect (with a 6.35  mm reference probe  and three  6.35  mm probes)
Operating conditions 0 °C to 35 °C, 0 % to 90 %
Environmental conditions for all specifications except temperature range 13 °C to 33 °C
Immersion (well)  depth 160  mm (6.3 in)
Well diameter 30 mm (1.18 in)
Heating time [1] –95 °C to 140 °C: 40 min
Cooling  time [1] 23 °C to –90 °C: 80 min
Stabilization time [2] 15 min
Resolution 0.01  °
Display LCD, °C or °F user selectable
Size (H x W x D) 480  mm x 205  mm x 380  mm (18.8 in x 8.0 in x 14.9  in)
Weight 16 kg (35 lb)
Power requirements 100 V to 115 V (± 10 %) 50/60 Hz, 575 W
System fuse  ratings 115 V: 6.3 A T 250 V
4–20 mA Fuse  (-P model only) 50 mA F 250 V
Computer interface RS-232, USB Serial, and 9930 interface-it temperature calibration software included
Safety IEC 61010-1, Installation Category II, Pollution degree 2
Electromagnetic environment IEC 61326-1: Basic
Refrigerants R32 (Difluoromethane)
Built-in reference thermometer readout accuracy (4-wire reference probe) [3] ± 0.010 °C at –95 °C
Reference resistance range 0 Ω to 400 Ω
Reference resistance accuracy [4] 0 Ω to 42 Ω: ± 0.0025 Ω   42 Ω to 400  Ω: ± 60 ppm of reading
Reference characterizations ITS-90, CVD, IEC-751, resistance
Reference measurement capability 4 wire
Reference probe connection 6-pin din with INFO-CON technology
Built-in RTD thermometer readout accuracy NI-120: ± 0.015 °C at 0 °C
RTD resistance range 0 Ω to 400 Ω
Resistance accuracy [4] 0 Ω to 25 Ω: ± 0.002 Ω
RTD characterizations PT-100 (385), (JIS), (3926), NI-120, resistance
RTD measurement capability 2-wire, 3-wire, and 4-wire RTD with jumpers  only
RTD connection 4-terminal input
Built-in TC thermometer readout accuracy [5] Type J: ± 0.70 °C at 140 °C
TC millivolt range –10  mV to 75 mV
Voltage accuracy 0.025 % of reading +0.01 mV
Internal cold  junction compensation accuracy ± 0.35 °C (ambient of 13 °C to 33 °C)
TC connection Miniature connectors (ASTM E1684)
Built-in mA readout accuracy 0.02  % of reading + 0.002 mA
mA range Cal 4-22 mA, Spec 4-24 mA
mA connection 2 terminal  input
Loop power function 24 V dc loop power
Built-in electronics temperature coefficient   (0 °C to 13 °C, 33 °C to 50 °C) ± 0.005 % of range  per °C