Additel ADT878PC-160-NO-110V – Reference Dry Well Calibrator, -40°C to 160°C, No Insert, 110V with Process Calibrator Option
$12,147.00
Technical Specs
Additel ADT878PC-160-NO-110V - Reference Dry Well Calibrator, -40°C to 160°C, No Insert, 110V with Process Calibrator Option Specs
| Temperature Range at 23°C | -40°C to 160°C |
| Display Accuracy | ±0.1°C at Full Range |
| Stability (30 min) | ±0.005°C at Full Range |
| Axial Uniformity at 60 mm (2.4 in) | ±0.025°C at -40°C ±0.020°C at 0°C ±0.050°C at 160°C |
| Axial Uniformity at 80 mm (3.15 in) | ±0.050°C at -40°C ±0.040°C at 0°C ±0.050°C at 160°C |
| Radial Uniformity | ±0.01°C at Full Range |
| Lading Effect | ±0.08°C (Display Sensor) ±0.010°C (External Sensor) |
| Hysteresis (Display Sensor) | 0.025°C |
| Environmental Conditions | 8°C to 38°C guaranteed accuracy 0°C to 50°C, 0% to 90% RH non-condensing |
| Storage Conditions | -20°C to 60°C |
| IP Rating | IP20 |
| Immersion Depth | 160 mm (1.30 in) |
| Insert OD | 31.9 mm (1.26 in) |
| Heating Time | 4 min: -40°C to 23°C 10 min: 23°C to 160°C |
| Cooling Time | 8 min: 160°C to 23°C 15 min: 23°C to -40°C |
| Typical Time to Stability | 10 min |
| Resolution | 0.001°C |
| Units | °C, °F, and K |
| Display | 6.5 in (165 mm) color touch screen |
| Size (H x W x D) | 170 x 345 x 330 mm (6.69 x 13.58 x 13.0 in) |
| Weight | 11.2 kg (24.7 lbs) |
| Power Requirements | 90-254 VAC, 45-65 Hz, 580 W |
| Communication | USB A, USB B, RJ45, WiFi, Bluetooth |
| Localization | English, Chinese, Japanese, Russian, German, French, Italian, and Spanish |
| Warranty | 1 yr |
| Readout Accuracy for 100 ohm PRT (Probe Accuracy Not Included) | ±0.006°C at 0°C ±0.008°C at 50°C ±0.009°C at 100°C ±0.011°C at 160°C ±0.015°C at 300°C ±0.019°C at 425°C ±0.026°C at 660°C ±0.028°C at 700°C |
| Readout Resolution | 0.1 mΩ |
| Reference Resistance Range | -200Ω to 926Ω |
| Reference Resistance Accuracy | 0Ω to 50Ω: ±1.25mΩ 50Ω to 400Ω: ±0.0025% RD |
| Reference Characterizations | ITS-90, CVD, IEC-751 |
| Reference Measurement Capability | 4-wire PRT |
| Reference Probe Connection | 6-pin lemo smart connector and Quick-Push connectors to accept banana, mini-banana, large & small spade lug and bare wire connections |
| RTD Channels | 2 |
| RTD Measurement Accuracy (excl sensor) Compliance | 0Ω to 50Ω: ±2.5mΩ 50Ω to 4000Ω: ±0.005% RD |
| RTD Measurement Resolution | 0.1mΩ |
| RTD Measurement Resistance Range | 0Ω to 4KΩ |
| RTD Characterizations | PT10, PT25, PT50, PT100, PT200, PT500, PT1000, CU10, CU50, CU100, NI100, NI120 |
| RTD Connection | Quick-Push connectors accept banana, mini-banana, large & small spade lug and bare wire connections |
| RTD Channels | 2 channels. Both accept 2, 3, or 4-wire RTDs |
| TC Channel | 2 |
| TC Measurement Channels | Accepting S, R, K, B, N, E, J, T, C, D, G, L, and U |
| TC Range | -75 mV to 75 mV |
| TC Resolution | 0.1µV |
| TC Voltage Accuracy | 0.01% RD + 5 µV |
| Internal CJC Accuracy | ±0.2°C (ambient from 0°C to 50°C) |
| Current Range | -30 mA to 30 mA |
| Current Accuracy | 0.01% RD + 2 µA |
| Current Resolution | 0.1 µA, Input Impedance: < 10Ω |
| Voltage Range | -30 V to 30 V |
| Voltage Accuracy | ±0.01% RD + 0.6 mV |
| Voltage Resolution | 0.1 mV; Input impedance: >1MΩ |
| Switch Test | Mechanical or Electrica |
| DC 24V Output | 24 V ±0.5 V, MAX 60 mA |
| Hart Communicator | Read, configure and calibrate HART devices - DD files updated periodically Optional - (order ADT875PC) |
| Documentation | Up to 1,000 tasks which store up to 10 results each containing as found and as left data. Snap shot feature allows for screen captures. Records auto step and ramp functions. |
| Temperature Coefficient 0℃ to 13℃ and 33℃ to 50℃ | ADT878 (PC)-160: ±0.005°C/°C ADT878 (PC)-425/700: ±0.005°C/°C Ref Readout: ±1 ppm FS/°C RTD Readouts: ±1 ppm FS/°C TC Readouts: ±5 ppm FS/°C Current: ±5 ppm FS/°C Voltage: ±5 ppm FS/°C |
| TC Type | Temperature (°C) |
| B | 250 |
| 300 | ±1.65 |
| 425 | ±1.18 |
| 660 | ±0.81 |
| 700 | ±0.77 |
| 1768 | ±0.56 |
| K | -200 |
| -40 | ±0.13 |
| 0 | ±0.13 |
| 160 | ±0.14 |
| 300 | ±0.15 |
| 425 | ±0.16 |
| 660 | ±0.18 |
| 700 | ±0.19 |
| 1000 | ±0.31 |
| E | -200 |
| -40 | ±0.09 |
| 0 | ±0.09 |
| 160 | ±0.08 |
| 300 | ±0.09 |
| 425 | ±0.10 |
| 660 | ±0.12 |
| 700 | ±0.13 |
| 1000 | ±0.17 |
| J | -210 |
| -40 | ±0.10 |
| 0 | ±0.10 |
| 160 | ±0.11 |
| 300 | ±0.12 |
| 425 | ±0.13 |
| 660 | ±0.14 |
| 700 | ±0.14 |
| 1000 | ±0.21 |
| T | -200 |
| -40 | ±0.14 |
| 0 | ±0.13 |
| 160 | ±0.11 |
| 300 | ±0.11 |
| 400 | ±0.11 |
| N | -200 |
| -40 | ±0.20 |
| 0 | ±0.19 |
| 160 | ±0.17 |
| 300 | ±0.17 |
| 425 | ±0.17 |
| 660 | ±0.19 |
| 700 | ±0.19 |
| 1000 | ±0.27 |
| S | -50 |
| -40 | ±1.17 |
| 0 | ±0.93 |
| 160 | ±0.63 |
| 300 | ±0.57 |
| 425 | ±0.55 |
| 660 | ±0.54 |
| 700 | ±0.53 |
| 1768 | ±0.66 |
| R | -50 |
| -40 | ±1.23 |
| 0 | ±0.95 |
| 160 | ±0.61 |
| 300 | ±0.54 |
| 425 | ±0.51 |
| 660 | ±0.48 |
| 750 | ±0.18 |
| 1768 | ±0.58 |
Reviews (0)
Be the first to review “Additel ADT878PC-160-NO-110V – Reference Dry Well Calibrator, -40°C to 160°C, No Insert, 110V with Process Calibrator Option” Cancel reply
You must be logged in to post a review.

Reviews