Technical Specs
Myron L 756II-116 Cond/TDS-Analog Monitor Only Specs
| Ranges: | 7 ranges from 0-200K to 0-20MΩ |
| Display: | Analog Models: 2.5 in./63 mm meter Digital Models: .5 in./13 mm 3½ Digit LCD Optional: 3½ Digit Backlit |
| Accuracy: | Analog Models: ± 2% of full scale Digital Models: ± 1% of full scale |
| Sensitivity: | 0.05% of full scale |
| Stability: | 0.05% of full scale |
| Repeatability: | 0.1% of full scale |
| Calibration Check: | Built-in |
| Recorder Output: | 0-10 VDC @ 5 mA max. (linear); standard all models |
| Sensor Inputs: | 1 (3 input option available) |
| Control Function: | Single Alarm/Controller fully adjustable 0-100% of full scale Hysteresis: Adjustable from 0.3-3% of full scale Indicators: Above (green) and below (red) setpoint LEDs (reversible) Contact Rating: SPDT 10 amp @ 250 VAC, 30 VDC. Relay operates decreasing or increasing reading (selectable) Second Alarm/controller, with above specifications - optional |
| Solid State Output: | Powered; 24 VDC unregulated 30 mA max |
| Power: | 115/230 VAC, 50/60 Hz, ±15% (user selectable) 24 VAC or VDC optional |
| Maximum Power Consumption: | @ 115V: 100 milliamps |
| Dimensions: | 152 H x 122 W x 96 D mm / 6 H x 4.8 W x 3.8 D in. |
| Construction: | Fully gasketed heavy-duty ABS case; water-resistant and corrosion proof IP65/NEMA 4X rated |
| Ambient Temp Range: | 0 - 60°C/32 - 140°F |
| Weight: | 0,9 kg/2 lbs. |